l
High UV-Vis performance between 200 and 1000 nm
l
Entry wavelength by keyboard
l
Single-point curve calibration and multi-point curve calibration (up to 12 points)
l
Display of fitting degree of curve fitting and concentration data with up to 50 sets of curve parameters storage
l
Display of concentration data with up to 200 sets of data storage
l
Concentration factor setting and concentration direct reading
l
Dark current calibration / wavelength calibration/system clock management
l
USB data communication
l
8 key keyboard
l
Internal 6 common working wavelength
Technique parameter
|
Optical System
|
Single beam, C-T type, 1200 lines/mm grating
|
Detector
|
Silicon photo diode
|
Monitor
|
20
×2LCD
|
Light Source
|
Tungsten lamp (socket type, 20W / 12V, 2000 hours)
Deuterium lamp (1000 hours)
|
Spectral Bandwidth
|
4nm
|
Wavelength
Range
|
200~1000nm
( step interval 0.1nm
)
|
Wavelength Accuracy
|
±2.0nm
( Auto-calibration when starting
)
|
Wavelength Repeatability
|
1.0nm
|
Stray Light
|
≤0.5%T
(at 220,340nm
)
|
Photometric Range
|
T
:-1.0~200.0%
,A
:-0.5~3.000Abs
,F
:0~9999
,C
:0~9999
|
Photometric Accuracy
|
±0.5%T
(0~100%T
)
|
Photometric Repeatability
|
0.2%T
|
Bright current
|
≤0.5%T
|
Dark current
|
≤0.2%T
|
Data Output
|
USB interface, LPT parallel printer interface
|
Power Requirement
|
220V±10% 50Hz
,150VA
|
Dimensions
|
455
×430
×220 mm
|
Gross Weight
|
22.5 kg
|
Net Weight
|
17.5kg
|
1x2arena.net