l
High Vis performance between 335 and 1000 nm
l
Specially designed for industry and labs
l
High accuracy with digital wavelength display
l
Automatic 0%(T) and 100%
(
T
)
calibration
l
Display of concentration data function
l
RS232 communication
l
Data-processing software (optional)
Technical parameter
|
Optical System
|
Single beam, C-T type, 1200 lines/mm grating
|
Detector
|
Silicon photo diode
|
Light Source
|
Tungsten lamp (socket type, 20W / 12V, 2000h)
|
Spectral Bandwidth
|
4nm
|
Wavelength
Range
|
Setting Wavelength
:332~1003nm
;
Measuring Wavelength
:335~1000nm
|
Wavelength Accuracy
|
±2nm
|
Wavelength Repeatability
|
1nm
|
Stray Light
|
≤0.5%T
(360nm
)
|
Photometric Range
|
T
:0~199.9%
,A
:-0.3~2.5Abs
,F
:0~9999
,C
:0~9999
|
Photometric Accuracy
|
±0.5%T
|
Photometric Repeatability
|
0.2%T
|
Stability
|
≤0.5%T/5min
|
Power Requirement
|
220V±10% 50Hz 45VA
|
Actual Dimensions
|
385×330×185
(mm
)
|
Net Weight
|
10kg
|
1x2arena.net
We sell cleanroom consumables , HPLC parts globally